Comparison · QB AutoSampler vs. Flownamics Seg-Flow S3 / PS
QB AutoSampler vs Flownamics Seg-Flow Comparison
Important notice. This comparison is published by A4BEE Sp. z o.o., the company behind the QB Systems brand and the QB AutoSampler, and therefore is not independent. It compares products intended for related purposes (automated sampling in laboratory- and pilot-scale bioprocessing) on the basis of publicly available information as of 19 October 2026, comparing the QB AutoSampler (QB Control sampling services) with the Flownamics Seg-Flow S3 / PS. It contains no prices and is not an offer. See the full Legal Notice & Disclaimer at the end of this document.
Same bench, same vessels · drag to swap the sampler
QB AutoSampler and Flownamics Seg-Flow S3 / PS on a laboratory bench
Bench top 2.1 m × 0.62 m, dotted grid every 500 mm. Lines show the overall width and depth of each set; QB shows its minimum and maximum width. Vessels are generic (Ø160 × 300 mm) and illustrative.
Dimensions and sources (W × H × D, mm)
- QB Selector Valve SVL-001 115 × 115 × 320 mm datasheet lists 320 × 115 × 115; drawn with the 320 mm side as depth
- QB QB Edge EDG-001 115 × 115 × 310 mm datasheet W × H × D
- QB Syringe Pump SPM-002 60 × 240 × 120 mm datasheet lists 240 × 60 × 120 without syringe; drawn upright
- QB Sample Cooling Chamber SCM-001 (optional) 237 × 273 × 232 mm optional, dashed; samples can also go straight into vials. Draft datasheet; 278 mm deep with connectors
- QB QB Thermo TCT-001 chiller (optional) 345 × 248 × 215 mm optional, dashed; axis order not stated, orientation indicative
- Flownamics Seg-Flow S3 317.5 × 508 × 356 mm 10.52 kg; published as 317.5 mm wide, 508 mm high, 356 mm deep
- Flownamics Seg-Flow PS 108 × 321 × 170 mm published as 12.64 in T × 6.68 in D × 4.24 in W; read as 321 mm tall, 170 mm deep, 108 mm wide and drawn upright
QB sizes are from the QB datasheets (the SCM-001 datasheet is a draft). Flownamics sizes are from the manufacturer's published specifications for the Seg-Flow S3 and the Seg-Flow PS; the PS axis reading (T as height) is our interpretation.
This page is the Flownamics extract of our longer article, The State of Benchtop Automated Sampling in Bioprocessing: A Comparative Analysis for 2026. Evidence labels in the QB column: C = QB Systems documentation and platform behaviour; K = QB product documentation 09.2026; O = statement of QB Systems (product owner), not verified in the product code. Flownamics values carry the source id from the register below. “Not stated” means the reviewed source does not say; it does not mean the feature is absent.
Choose QB AutoSampler when
- Cooled sample storage in the QB Sample Cooling Chamber for later analysis
- Recipe-driven triggers (timers, counters, live device tags) on one control platform
- Non-regulated research and process development
- Optional in-vial dilution and aliquoting on the same hardware
Choose Flownamics Seg-Flow S3 / PS when
- Feeding at-line analysers (for example a Vi-CELL BLU) with in-line dilution
- A documented cleaning step after every sample
- Sampling from up to eight vessels or process streams to up to four analysers
- Cell and gene therapy workflows (Seg-Flow PS)
New to automated sampling? Our guide to automated aseptic bioreactor sampling explains how samplers work and how to choose one. Full details of the QB system are on the QB AutoSampler page.
Side by side
Sampling points
Selector-valve ports; the valve is stackable
1 to 8 vessels or process streams
Destinations
Vials in the QB Sample Cooling Chamber holders or other destination ports
Up to four analysers and/or fraction collectors
Cooled storage
4 °C for more than 72 h while the cooling service runs
Temperature-controlled FlowFrac 1000 fraction collector
Sample volume
Recommended 1.0 to 6 mL; dead volume 0.5 to 1.0 mL
Not stated
Cleaning
Configurable wash, rinse and air-dry legs; port-level Sampler Cleaning
Central Reagent-Manifold; line cleaned and/or sterilised after every sample
Dilution
In-vial, factor 1.0 to 6.0
Optional Sample-Mod S3
How each sample path is built
Sources
Core
Destinations
Schematic of the documented path, not to scale.
Sources
Core
Destinations
Schematic of the documented path, not to scale.
At a glance
- Different architectures. The Seg-Flow S3 is a segmented-flow sampler that delivers to analysers or fraction collectors and cleans the sample line after every sample. The QB AutoSampler is a syringe-pump and selector-valve sampler on the QB Control platform, with in-vial dilution and optional cooled storage.
- Not stated by Flownamics: commercial terms and licence terms [S3, S4]. This page compares no prices.
1. Sources and modularity
| QB AutoSampler | Flownamics Seg-Flow S3 / PS | |
|---|---|---|
| Architecture | QB Syringe Pump (SPM-002), QB Selector Valve (SVL-001, up to 24 positions shared between source, wash, air, waste and destination ports), QB Edge (EDG-001) and QB Control services with a recipe engine [C, K] | Segmented-flow sampling to analysers or fraction collectors [S3] |
| Sampling points | Selector-valve ports; the valve is stackable [O] | Samples from 1 to 8 vessels or process streams [S3] |
| Modularity | Base services Automated Sampling, Sampler Cleaning and Temperature Control by TCU; Sampling Dilution and Sampling Aliquoting are optional services on the same hardware [C] | Optional Sample-Mod S3 for dilution; temperature-controlled FlowFrac 1000 fraction collector [S3] |
| Seg-Flow PS | n/a | For cell and gene therapy; “One (1) per PS, up to 16 PS in parallel” in the specification table [S4] |
2. Destinations
| QB AutoSampler | Flownamics Seg-Flow S3 / PS | |
|---|---|---|
| Destinations | Vials in the QB Sample Cooling Chamber holders or other destination ports on the selector valve [C, K] | Up to four analysers and/or fraction collectors [S3] |
| Analyser integration | No specific analyser integration and no maximum number of analysers documented [C] | Joint page of Beckman Coulter for the Vi-CELL BLU and Seg-Flow S3 states load-and-go runs for up to eight bioreactors [S6] |
| Seg-Flow PS | n/a | Moves samples up to 50 ft; with the Director, up to four analytical devices [S4] |
3. Sample storage
| QB AutoSampler | Flownamics Seg-Flow S3 / PS | |
|---|---|---|
| Cooled storage | QB Sample Cooling Chamber (SCM-001) with a QB Thermo chiller (TCT-001, 2 to 42 °C) and a QB Multisensor (MTS-001) [K]; 4 °C for more than 72 h [O]; cooling only while the Temperature Control by TCU service is in Execute [C] | Temperature-controlled FlowFrac 1000 fraction collector available; temperature range not stated [S3] |
| Holders | 42 x 4 mL, 36 x 8 mL or 21 + 18 mixed [K] | Not stated [S3] |
| Cool-down | Roughly 10 min to 4 °C in a pre-cooled chamber, roughly 20 min from room temperature, 1 °C minimum in lab conditions, as stated in the QB catalog [K] | Not stated [S3] |
4. Sample volume and cleaning
| QB AutoSampler | Flownamics Seg-Flow S3 / PS | |
|---|---|---|
| Sample volume | Recommended 1.0 to 6 mL (6 mL needs the 10 mL syringe) [O]; dead volume 0.5 to 1.0 mL [O]; path accuracy approx. ±0.5 mL [O]; pump accuracy ±0.5 % [K] | Not stated [S3] |
| Cleaning | Configurable wash, rinse and air-dry legs plus optional path fill; “wash after sample” is on by default; Sampler Cleaning washes selected valve ports [C] | A central Reagent-Manifold cleans and disinfects the fluid path, and the sample line is cleaned and/or sterilised after every sample (vendor statement) [S3] |
| Fault detection | See limits below | Redesigned unit with self-testing intelligent pinch valves and non-contact ultrasonic sensors that flag blocked or misconnected tubing, an improved peristaltic pump, a built-in Reagent Module, a live flow-path display and status LEDs; the redesign date is not stated [S3] |
5. Dilution and splitting
| QB AutoSampler | Flownamics Seg-Flow S3 / PS | |
|---|---|---|
| Dilution | Sampling Dilution: in-vial, factor 1.0 to 6.0, air chase, no mixing, no specified ratio accuracy, sample of at least 1 mL when D is above 1, one destination per run [C] | Optional Sample-Mod S3: pre-dilution by reagent addition and serial dilution; Flownamics states no ratio [S3]. Beckman Coulter states in-line dilutions “up to 1:20” for the joint Vi-CELL BLU and Seg-Flow S3 solution [S6] |
| Splitting | Sampling Aliquoting: one draw split sequentially to two ports, no numeric accuracy specified [C] | Distribution to up to four analysers and/or fraction collectors [S3] |
6. Triggers and control
| QB AutoSampler | Flownamics Seg-Flow S3 / PS | |
|---|---|---|
| Triggers | Recipes with live device tags, timers, counters, variables and signals; external trigger through the OPC UA connector [O] (no sampling-specific mapping found in the code); status events for external clients [C] | Not stated [S3] |
| Data interfaces | OPC UA server [C] | FlowWeb exports data to OPC-enabled SCADA, with an optional OPC server package [S3] |
| Feedback control | No analyser-result feedback control documented [C] | Not stated [S3] |
7. Data integrity and licensing
| QB AutoSampler | Flownamics Seg-Flow S3 / PS | |
|---|---|---|
| Records | Audit trail, PDF batch reports and a “Sampling basic” report [C] | Not stated [S3] |
| 21 CFR Part 11 | Conformance features for the requirements it implements [O]; electronic signatures not yet covered (in development, no committed date). QB Systems does not claim compliance or validation | A 2021 Flownamics brochure (third-party hosted, last checked 10 October 2026) makes a 21 CFR Part 11 claim [S5]. The current product page makes no Part 11 statement [S3]; confirm the current status with the vendor. The Beckman Coulter page says “GMP compliance with electronic data management” for the joint solution [S6] |
| Licensing | Tiers BASIC, ADVANCED and PRO, with validity dates and expiry warnings and per-service licence entries [C]; 12 months of support included [O] | Not stated [S3, S4] |
Where the QB AutoSampler has limits
- Dilution is in-vial only and limited to a factor of 6; splitting goes sequentially to two ports. Securecell Numera and Flownamics Seg-Flow S3 publish distribution to up to six and four analysers respectively [S1, S3].
- No specific analyser integration and no maximum number of analysers is documented; the Seg-Flow S3 documents up to four [S3].
- Dead volume and path accuracy are disclosed; comparable vendor figures were not found.
- Known limitation: when the second-aspiration volume plus the sample exceeds the syringe volume, a smaller sample is taken without a warning.
- Cooling is available only while the Temperature Control by TCU service runs, so cooling continuity needs planning.
- Sampling Aliquoting r1 has no process limits and no six-level alarms.
- No published list of verified bioreactor interfaces; compatibility with the Flownamics FISP probe has not been tested [O].
- The QB AutoSampler is positioned for non-regulated research and process development. Electronic signatures are not yet covered (in development, no committed date).
- Newer entrant: no application data is cited here.
- No LIMS / ELN integration is documented.
Next step
Request a quote from each vendor: Flownamics commercial and licence terms are not stated in the sources reviewed.
Sources
Flownamics and QB sources only; the full register is in the article.
- S1 Securecell, Numera product information, as listed in the article register (cited for the limits only)
- S3 Flownamics, Seg-Flow S3 product page, https://flownamics.com/product/automated-sampling-system/, 19 Oct 2026
- S4 Flownamics, Seg-Flow PS product page, https://flownamics.com/product/seg-flow-ps/, 19 Oct 2026
- S5 Flownamics, Seg-Flow brochure (2021, hosted by a third party; may be outdated), https://making.com/wp-content/uploads/2021/06/Brochure_Flownamics-Seg-Flow.pdf, 10 Oct 2026 (not re-checked)
- S6 Beckman Coulter, Vi-CELL BLU and Seg-Flow S3, https://www.beckman.com/cell-counters-and-analyzers/vi-cell-blu/flownamics, 19 Oct 2026
- QB-C QB Systems documentation and platform behaviour; state of 14 October 2026
- QB-K QB product documentation 09.2026
- QB-O Statements of QB Systems (product owner), not verified in the product code
Legal Notice & Disclaimer
1. Publisher and transparency. This document is prepared and published by A4BEE Sp. z o.o. (Poland), which develops and sells QB Control and QB Modules under the QB Systems brand. A4BEE has a commercial interest in the products compared. The comparison is intended to highlight objective differences; it is not an independent test or review.
2. Nature of the comparison. This document is comparative information within the meaning of Directive 2006/114/EC. It compares products that meet related needs or are intended for a related purpose, and addresses material, relevant, verifiable and representative features. It is not intended to discredit or denigrate any company, product or trademark, to create confusion between brands, or to take unfair advantage of the reputation of any trademark.
3. Basis, date and versions. The information on third-party products is based on publicly available sources listed under Sources, gathered on or before 19 October 2026 (the 2021 Flownamics brochure on 10 October 2026), and refers to the Flownamics Seg-Flow S3 / PS. QB AutoSampler information refers to the sampling services of QB Control (QB Systems documentation of 14 October 2026). Products, features, versions, licensing terms and documentation change over time. Where a feature is described as “not stated”, this means it was not found in the public sources reviewed; it does not mean that the feature does not exist. Readers should verify current information directly with each vendor before making any decision.
4. Methodology. This page contains no ratings, scores or “winner” labels. Statements are A4BEE’s reading of the sources listed, labelled in the evidence column. Different use cases or weightings may lead to different conclusions.
5. No prices, no offer, no warranty. This document intentionally contains no prices. Nothing in it is an offer, quotation, specification or contractual commitment, and nothing in it constitutes a warranty of fitness for a particular purpose. QB Control capabilities depend on the software version, licence tier (BASIC / ADVANCED / PRO), configuration and connected hardware. Binding terms apply only under a signed agreement with A4BEE Sp. z o.o.
6. Regulatory statements. Statements about FDA 21 CFR Part 11, EU GMP Annex 11, GAMP 5 or other standards describe product features only. They are not certifications, legal or regulatory advice, or a guarantee of compliance. Compliance, validation and the GAMP categorisation of any computerised system remain the responsibility of the regulated user. QB Control provides 21 CFR Part 11 conformance features (electronic records, audit trail, access control, user management); electronic signatures are not yet covered (in development, no committed date).
7. Forward-looking statements. Any reference to future development is a statement of current intent only, not a commitment, and may change without notice.
8. Trademarks. All trademarks, logos and brand names are the property of their respective owners. They are used only to identify the products compared (nominative use). Their use does not imply any endorsement by, affiliation with, or sponsorship from their owners. QB Systems, QB Control, QB AutoSampler and QB Modules are brand names of A4BEE Sp. z o.o.
9. Limitation of liability. This document is provided “as is” for general information. To the maximum extent permitted by applicable law, A4BEE Sp. z o.o. accepts no liability for errors or omissions, or for any loss or damage arising from reliance on this document or from decisions made on its basis.
10. Corrections. If you believe any statement about your product is inaccurate or out of date, please contact [email protected] with supporting information. A4BEE will review the request promptly and correct or remove inaccurate content.
11. Copyright. © 2026 A4BEE Sp. z o.o. This document may be shared only in full, unmodified, and together with this notice.
Disclaimer: Competitor information on this page comes from the public vendor pages and third-party pages listed in the sources, retrieved on 19 October 2026 unless the sources state an earlier date. Vendors may change specifications at any time; “not stated” means the reviewed source does not say, not that a feature is absent. The page is not exhaustive and may omit options or configurations. Confirm all specifications and commercial terms directly with each vendor. QB Systems supplied the specifications of the QB AutoSampler (QB Systems documentation and statements of QB Systems, labelled in the evidence column); competitor data was collected independently from the sources cited. Corrections from vendors are welcome at [email protected]. The QB AutoSampler is intended for non-regulated research and process development; electronic signatures are not yet covered (in development, no committed date).