Comparison · QB AutoSampler vs. Merck MAST
QB AutoSampler vs Merck MAST Comparison
Important notice. This comparison is published by A4BEE Sp. z o.o., the company behind the QB Systems brand, QB Control and the QB AutoSampler, and therefore is not independent. It compares products intended for the same purpose (automated aseptic sampling of bioreactors) on the basis of publicly available information as of 19 October 2026, comparing the QB AutoSampler (QB Control sampling services) with Merck MAST. It contains no prices and is not an offer. Statements about QB are labelled by evidence type. See the full Legal Notice & Disclaimer at the end of this document.
Same bench, same vessels · drag to swap the sampler
QB AutoSampler and Merck MAST on a laboratory bench
Bench top 2.1 m × 0.62 m, dotted grid every 500 mm. Lines show the overall width and depth of each set; QB shows its minimum and maximum width. Vessels are generic (Ø160 × 300 mm) and illustrative.
Dimensions and sources (W × H × D, mm)
- QB Selector Valve SVL-001 115 × 115 × 320 mm datasheet lists 320 × 115 × 115; drawn with the 320 mm side as depth
- QB QB Edge EDG-001 115 × 115 × 310 mm datasheet W × H × D
- QB Syringe Pump SPM-002 60 × 240 × 120 mm datasheet lists 240 × 60 × 120 without syringe; drawn upright
- QB Sample Cooling Chamber SCM-001 (optional) 237 × 273 × 232 mm optional, dashed; samples can also go straight into vials. Draft datasheet; 278 mm deep with connectors
- QB QB Thermo TCT-001 chiller (optional) 345 × 248 × 215 mm optional, dashed; axis order not stated, orientation indicative
- Merck MAST Controller 394 × 175 × 351 mm datasheet size, includes clearance
- Merck MAST Sample Navigator 213 × 109 × 218 mm datasheet size, includes clearance
- Merck MAST Sample Pilot 100 114 × 91 × 100 mm datasheet size, includes clearance
- Merck MAST Sample Pilot 200 61 × 79 × 48 mm datasheet size, includes clearance
- Merck MAST Cleaning System 546 × 508 × 386 mm datasheet size, includes clearance
QB sizes are from the QB datasheets (the SCM-001 datasheet is a draft). Merck MAST sizes are from the MAST datasheet (ds12917en) and include clearance, so the devices themselves are somewhat smaller than drawn. A MAST installation uses the units it needs; all five are shown for scale.
Choose QB AutoSampler when
- Modular, software-defined benchtop sampling path with recipe-driven triggering
- Cooled vial storage at 4 °C while the temperature control service runs
- In-vial dilution (factor up to 6) and split aliquoting between two ports
- Audit trail, PDF batch reports and a Sampling basic report
Choose Merck MAST when
- Documented integration with supervisory systems such as DeltaV and PI
- Documented downstream liquid handler integration (Gilson)
- Up to four primary analytical destinations plus secondary destinations
- Sample Pilots for stainless steel (SP100) and development-scale or single-use (SP200) bioreactors
New to automated sampling? Our guide to automated aseptic bioreactor sampling explains how samplers work and how to choose one. Full details of the QB system are on the QB AutoSampler page.
Side by side
Architecture
Syringe pump and selector valve on a QB Edge; scale by adding valve and pump pairs
Sample Pilots (SP100, SP200); Navigator Controller routes up to 5 pilots
Destinations
Vials, analytical devices and waste; no specific analyser integration documented
Up to four primary analytical destinations plus secondary destinations
Sample volume
Recommended 1.0 to 6 mL
SP100: 10 or 55 mL; SP200: 1 or 5 mL
Cooled storage
4 °C for more than 72 h while the cooling service runs
Not stated; a Gilson liquid handler integration is reported
Cleaning
Configurable wash, rinse and air-dry legs
Pre- and post-sample sanitation and cleaning
How each sample path is built
Sources
Core
Destinations
Schematic of the documented path, not to scale.
Sources
Core
Destinations
Schematic of the documented path, not to scale.
Merck MAST information comes from a data sheet dated April 2024 (version 1.0) and a technical article. It is not a recent source. “Not stated” means the reviewed source does not say, not that a feature is absent.
Evidence labels
C = QB Systems documentation and platform behaviour (state of 14 October 2026); K = QB product documentation 09.2026; O = statement of QB Systems (product owner), not verified in code; M = competitor manufacturer source. Merck sources: S8 (MAST Autosampling Solution data sheet ds12917en, version 1.0, April 2024) and S9 (MAST technical article).
Architecture and modularity
| System | What the source states | Source | Evidence |
|---|---|---|---|
| Merck MAST | Sample Pilots (SP100, SP200) with controllers; the Sample Navigator Controller routes up to 5 Sample Pilots; software configures up to 10 locations | S8 | M |
| QB AutoSampler | A QB Syringe Pump (SPM-002) and a QB Selector Valve (SVL-001) form a sampling path; one valve offers up to 24 positions shared between source, wash, air, waste and destination ports; valves are stackable; a QB Edge (EDG-001) powers and controls the modules over PoE Ethernet (seven PoE ports, more via QB Switch); paths can be shared, dedicated or combined; scaling by adding valve and pump pairs plus a licence upgrade | QB | C (port roles); K (QB Edge); O (stackable valves, scaling) |
Valve positions are not sources, so a number of positions is not compared with a number of Sample Pilots.
Destinations
| System | What the source states | Source | Evidence |
|---|---|---|---|
| Merck MAST | Up to four primary analytical destinations (named: Gilson GX-271, Nova FLEX2) plus multiple secondary destinations | S8 | M |
| QB AutoSampler | Vials, analytical devices and waste are destination types; the number of destinations is limited by the valve positions; no specific analyser integration and no maximum number of analysers are documented | QB | C (port roles); O (destination types) |
Cooled storage and downstream preparation
| System | What the source states | Source | Evidence |
|---|---|---|---|
| Merck MAST | A Gilson liquid handler integration is reported in the data sheet (chilled retain; not re-checked on 19 October 2026); storage conditions are not stated | S8 | M |
| QB AutoSampler | QB Sample Cooling Chamber SCM-001 with an external chiller loop: 4 °C for more than 72 h (only while the Temperature Control by TCU service is in Execute); 1 °C minimum reached in laboratory conditions; roughly 10 min to 4 °C in a pre-cooled chamber and roughly 20 min from a room-temperature chamber; 42 × 4 mL, 36 × 8 mL or 21 + 18 mixed vials per chamber; a 4 mL vial holds a maximum of 3 mL, an 8 mL vial a maximum of 6 mL | QB | K (holders, cooling times, 1 °C minimum); C (TCU service behaviour); O (more than 72 h at 4 °C, vial figures) |
Sample volumes and cleaning
| System | Sample volume | Cleaning | Source | Evidence |
|---|---|---|---|---|
| Merck MAST | SP100: 10 or 55 mL (stainless steel bioreactors); SP200: 1 or 5 mL (development-scale and single-use bioreactors) | Pre- and post-sample sanitation (70 % isopropyl alcohol) and cleaning (0.1–1 N NaOH); a sanitation and drying step clears the lines | S8 | M |
| QB AutoSampler | 1.0–6 mL recommended (6 mL needs the 10 mL syringe); dead volume 0.5–1.0 mL; path accuracy approx. ±0.5 mL | Configurable wash (N), rinse (N) and air-dry legs plus optional path fill (“wash after sample” on by default); a port-level Sampler Cleaning service; dedicated consumable sets, an A4BEE-provided SOP and an aseptic testing report on request (owner statement) | QB | C (service behaviour, cleaning); O (volume range, dead volume, path accuracy, aseptic set and SOP) |
Comparable dead-volume and path-accuracy figures were not found in the Merck sources, so no conclusion is drawn either way. QB Systems makes no claim about the contamination-control level of the AutoSampler tubing, pass-throughs or chamber.
Sample preparation
| System | What the source states | Source | Evidence |
|---|---|---|---|
| Merck MAST | Not stated for the sampler; a Gilson liquid handler integration is reported in the data sheet (dilutions and aliquots are named there; not re-checked on 19 October 2026); the technical article confirms the Gilson Verity GX-271 for sample preparation | S8, S9 | M |
| QB AutoSampler | Sampling Dilution Service: in-vial, dilution factor 1.0–6.0 (total divided by sample), no mixing (the vial must be vortexed), ratio accuracy not specified. Sampling Aliquoting Service: one draw split sequentially between two ports, no numeric accuracy specified, no process limits specified | QB | C (QB Systems documentation) |
Triggering and integration
| System | What the source states | Source | Evidence |
|---|---|---|---|
| Merck MAST | Interfaces with supervisory systems including DeltaV and PI | S8 | M |
| QB AutoSampler | Scheduled and recipe-conditional sampling using live device tags, timers, counters and variables; an OPC UA server for external clients; services publish status events for external clients; no feedback control from analyser results documented | QB | C (recipes, OPC UA server, MQTT) |
Data integrity and licensing
| System | What the source states | Source | Evidence |
|---|---|---|---|
| Merck MAST | GMP and 21 CFR Part 11 statements are not stated; the technical article states that each sample is logged with origin, purpose and operator metadata; commercial and licence terms are not stated | S8, S9 | M |
| QB AutoSampler | Audit trail (including Batch Run and Recipe Audit Trail reports, recipe import and export, safety actions, service outcomes), PDF batch reports and a “Sampling basic” report. QB Control provides 21 CFR Part 11 conformance features; electronic signatures are not yet covered (in development, no committed date). Licence tiers BASIC / ADVANCED / PRO, perpetual or fixed-term, with validity dates and expiry warnings; confirm the licence term with QB Systems | QB | C (audit trail, reports, tiers); O (conformance wording) |
Where the QB AutoSampler has limits
- Dilution is in-vial only, up to a factor of 6, without mixing and without a specified ratio accuracy.
- Analyser transfer. Analytical devices can be allocated as destination ports, but QB documents no specific analyser integration and no maximum number of analysers; Merck names its primary destinations.
- Cooling continuity. Cooling runs only while the Temperature Control by TCU service is in Execute; it stops when the service is paused, held, stopped, completed or aborted, when the message bus connection is lost and when the service restarts.
- Bioreactor connection. QB Systems has not published a list of verified bioreactor interfaces; confirm port compatibility per case.
- Supervisory integration. The reviewed QB sources document no DeltaV or PI integration.
- Integrations. No LIMS / ELN integration is documented.
- Positioning. The AutoSampler is positioned as non-regulated R&D and process-development tooling. QB Systems is a newer entrant in dedicated bioprocess sampling, and this comparison cites no application data for the AutoSampler.
Choosing between them
Merck MAST suits sites that want sampling integrated with supervisory systems such as DeltaV or PI and a downstream liquid handler. The QB AutoSampler suits labs that want a modular, software-defined sampling path with recipe-driven triggering, cooled vial storage and an audit trail on open protocols. Confirm all specifications with each vendor. For the full five-vendor analysis see The State of Benchtop Automated Sampling in Bioprocessing.
Legal Notice & Disclaimer
1. Publisher and transparency. This document is prepared and published by A4BEE Sp. z o.o. (Poland), which develops and sells QB Control and QB Modules, including the QB AutoSampler, under the QB Systems brand. A4BEE has a commercial interest in the products compared. The comparison is intended to highlight objective differences; it is not an independent test or review.
2. Nature of the comparison. This document is comparative information within the meaning of Directive 2006/114/EC. It compares products that meet the same needs or are intended for the same purpose, and addresses material, relevant, verifiable and representative features. It is not intended to discredit or denigrate any company, product or trademark, to create confusion between brands, or to take unfair advantage of the reputation of any trademark.
3. Basis, date and versions. The information on third-party products is based on publicly available sources listed under Sources, gathered on or before 19 October 2026, and refers to Merck MAST (MAST Autosampling Solution data sheet version 1.0, dated April 2024). QB information refers to the QB AutoSampler with QB Control and QB Systems documentation 09.2026. Products, features, versions, licensing terms and documentation change over time. Where a feature is described as “not stated”, this means it was not found in the public sources reviewed; it does not mean that the feature does not exist. Readers should verify current information directly with each vendor before making any decision.
4. Evaluation and methodology. Descriptions and the “Choosing between them” section are A4BEE’s qualitative assessment for laboratory-scale bioprocess sampling. They are not measurements, certifications or test results. Different use cases or weightings may lead to different conclusions. Statements about QB marked O are statements of QB Systems and have not been verified in the product code.
5. No prices, no offer, no warranty. This document intentionally contains no prices. Nothing in it is an offer, quotation, specification or contractual commitment, and nothing in it constitutes a warranty of fitness for a particular purpose. QB capabilities depend on the software version, licence tier (BASIC / ADVANCED / PRO), configuration and connected hardware. Binding terms apply only under a signed agreement with A4BEE Sp. z o.o.
6. Regulatory statements. Statements about FDA 21 CFR Part 11, EU GMP Annex 11, GAMP 5 or other standards describe product features only. They are not certifications, legal or regulatory advice, or a guarantee of compliance. Compliance, validation and the GAMP categorisation of any computerised system remain the responsibility of the regulated user. QB Control provides 21 CFR Part 11 conformance features (electronic records, audit trail, access control, user management); electronic signatures are not yet covered.
7. Forward-looking statements. Any reference to future development is a statement of current intent only, not a commitment, and may change without notice.
8. Trademarks. All trademarks, logos and brand names are the property of their respective owners. They are used only to identify the products compared (nominative use). Their use does not imply any endorsement by, affiliation with, or sponsorship from their owners. Trademark symbols follow the owners’ own usage as far as known. QB Systems, QB Control, QB Modules and QB AutoSampler are brand names of A4BEE Sp. z o.o.
9. Limitation of liability. This document is provided “as is” for general information. To the maximum extent permitted by applicable law, A4BEE Sp. z o.o. accepts no liability for errors or omissions, or for any loss or damage arising from reliance on this document or from decisions made on its basis.
10. Corrections. If you believe any statement about your product is inaccurate or out of date, please contact [email protected] with supporting information. A4BEE will review the request promptly and correct or remove inaccurate content.
11. Copyright. © 2026 A4BEE Sp. z o.o. This document may be shared only in full, unmodified, and together with this notice.
Sources
Competitor information comes from the public sources below, retrieved on 19 October 2026. The Merck MAST data sheet is dated April 2024. Vendors may change specifications at any time. The comparison is not exhaustive and may omit options or configurations. QB Systems supplied the specifications of the QB AutoSampler (QB Systems documentation and statements of QB Systems, labelled in the Evidence column).
- S8 — Merck, MAST Autosampling Solution data sheet ds12917en (version 1.0, April 2024)
- S9 — Merck, MAST technical article
- QB-C — QB Systems documentation and platform behaviour; state of 14 October 2026
- QB-K — QB product documentation 09.2026
- QB-O — statements of QB Systems (product owner), not verified in the product code
Merck®, MAST, Sigma-Aldrich®, Gilson®, Nova®, DeltaV™ and PI are trademarks of their respective owners.