Comparison · QB AutoSampler vs. Securecell Numera

QB AutoSampler vs Securecell Numera Comparison

Important notice. This comparison is published by A4BEE Sp. z o.o., the company behind the QB Systems brand and the QB AutoSampler, and therefore is not independent. It compares products intended for the same purpose (benchtop automated sampling of bioreactors) on the basis of publicly available information as of 19 October 2026, comparing the QB AutoSampler (QB Control sampling services) with the Securecell Numera. It contains no prices and is not an offer. See the full Legal Notice & Disclaimer at the end of this document.

Same bench, same vessels · drag to swap the sampler

QB AutoSampler and Securecell Numera 3 on a laboratory bench

182 mm min1175 mm max 320 mm
970 mm 540 mm
QB AutoSampler Numera 3

Bench top 2.1 m × 0.62 m, dotted grid every 500 mm. Lines show the overall width and depth of each set; QB shows its minimum and maximum width. Vessels are generic (Ø160 × 300 mm) and illustrative.

Dimensions and sources (W × H × D, mm)
  • QB Selector Valve SVL-001 115 × 115 × 320 mm datasheet lists 320 × 115 × 115; drawn with the 320 mm side as depth
  • QB QB Edge EDG-001 115 × 115 × 310 mm datasheet W × H × D
  • QB Syringe Pump SPM-002 60 × 240 × 120 mm datasheet lists 240 × 60 × 120 without syringe; drawn upright
  • QB Sample Cooling Chamber SCM-001 (optional) 237 × 273 × 232 mm optional, dashed; samples can also go straight into vials. Draft datasheet; 278 mm deep with connectors
  • QB QB Thermo TCT-001 chiller (optional) 345 × 248 × 215 mm optional, dashed; axis order not stated, orientation indicative
  • Numera 3 Module stack (example) 460 × 765 × 440 mm two 230 mm modules per row, three rows
  • Numera 3 Sample Collector ASX-7200 370 × 460 × 540 mm excludes syringe module, Peltier racks, injection valve

QB sizes are from the QB datasheets (the SCM-001 datasheet is a draft). Numera sizes are from the Numera 3 brochure (undated, reseller-hosted copy) and exclude the items noted. The Numera Peltier racks are inside the collector.

Isometric line art at one scale, redrawn by QB Systems. The bench, vessels and arrangement are an example, the units are simplified, and tubing and cabling are not shown. The QB AutoSampler is drawn with its selector valve, QB Edge and syringe pump; the sample cooling chamber and the QB Thermo chiller are optional (dashed), as samples can also go straight into vials, so the QB width is shown as a minimum and a maximum. The Numera 3 collector excludes its syringe module (dashed), Peltier racks and injection valve. Overall size is one dimension only and does not compare capability. Not a vendor illustration.

Choose QB AutoSampler when

  • A software-defined sampling path with configurable wash, rinse and air-dry legs, driven by recipes
  • Cooled vial storage (4 °C) integrated in the same control platform as the bioreactor control
  • Sampling triggered by live device tags, timers, counters or an external OPC UA client
  • Audit trail and PDF batch reports, including a Sampling basic report, in non-regulated R&D and process development
  • Starting small and growing by adding valve and pump pairs and a licence upgrade

Choose Securecell Numera when

  • Dilution over a wide range (1:2 to 1:100) and membrane filtration inside the sampler
  • Direct transfer to several analysers (up to six) with a named list of supported analysers
  • Sampling from up to 16 bioreactors with intervals as short as 6 minutes
  • Direct HPLC injection valves and an Ambr 250 interface
  • Feedback control from analyser results through Lucullus software

New to automated sampling? Our guide to automated aseptic bioreactor sampling explains how samplers work and how to choose one. Full details of the QB system are on the QB AutoSampler page.

Side by side

QB AutoSampler
Securecell Numera

Sources

Selector valve, up to 24 positions shared between source, wash, air, waste and destination ports

Up to 16 bioreactors

Analyser transfer

Not documented: no specific analyser integration

Direct transfer to up to six analysers; named analyser list

Dilution

In-vial, factor 1.0 to 6.0, no mixing

1:2 to 1:100 (precision not stated)

Filtration

Not documented

Membrane-based, 0.22, 0.45 and 1.20 µm tapes

Sample storage

Cooled chamber, 4 °C for more than 72 h while the cooling service runs

Peltier racks, 4 to 20 °C (product page) or 4 to 40 °C (catalogue)

Triggers

Scheduled and recipe-conditional; OPC UA

Interval as short as 6 minutes

How each sample path is built

QB AutoSampler

Sources

Selector-valve ports (up to 24 positions, shared)

Core

QB Syringe Pump (SPM-002)
QB Selector Valve (SVL-001)
QB Edge (EDG-001)
QB Control recipes

Destinations

Vials in the QB Sample Cooling Chamber
Analytical devices (destination ports)
Waste

Schematic of the documented path, not to scale.

Securecell Numera

Sources

Up to 16 bioreactors
Up to four Multiplexer Modules

Core

Control Module
Multiplexer and Routing Modules
Dilution Module (1:2 to 1:100)
Filtration Module (0.22, 0.45, 1.20 µm)

Destinations

Up to six analysers (named list)
HPLC injection valves
Sample Collector (ASX-7200, ASX-7400)

Schematic of the documented path, not to scale.

QB Syringe Pump
QB Syringe Pump (SPM-002)
QB Selector Valve with tubing connected to its ports
QB Selector Valve (SVL-001), up to 24 positions
QB Sample Cooling Chamber next to a vial holder
QB Sample Cooling Chamber (SCM-001) with vial holder

At a glance

Dimension QB AutoSampler Securecell Numera
Architecture Modular, software-defined: syringe pump, selector valve and QB Edge, services and recipes in QB Control Modular flow-based system: Control, Multiplexer, Routing, Dilution and Filtration Modules plus a separate Sample Collector
Sources Selector valve with up to 24 positions shared between source, wash, air, waste and destination ports; no fixed vessel maximum documented ◆ Up to 16 bioreactors (up to four Multiplexer Modules)
Analyser transfer Analytical devices can be destination ports; no specific analyser integration documented ◆ Direct transfer to up to six analysers; named analyser list; HPLC injection valves
Dilution In-vial, factor 1.0 to 6.0, no mixing, ratio accuracy not specified ◆ Dilution Module, 1:2 to 1:100 (precision not stated; ratio notation not directly comparable)
Filtration Not documented ◆ Membrane-based tangential filtration (0.22, 0.45 and 1.20 µm tapes)
Sample storage Cooled chamber with chiller loop, 4 °C for more than 72 h while the cooling service runs Peltier racks for vials and deep-well plates, 4 to 20 °C (product page) or 4 to 40 °C (catalogue)
Triggers and control Scheduled and recipe-conditional sampling; OPC UA server; status events for external clients ◆ Interval as short as 6 minutes; coordinated by Lucullus software incl. feedback control
Data integrity Audit trail and PDF batch reports; Part 11 conformance features; electronic signatures not yet covered (in development, no committed date) Part 11 statements not stated in the sources reviewed
Licensing BASIC / ADVANCED / PRO tiers, perpetual or fixed-term; scaling by licence upgrade Licence terms not stated in the sources reviewed

What each system is

Securecell Numera is a modular flow-based sampling system, now branded “Numera 3”. It consists of a Control Module, a Multiplexer Module, a Routing Module, a Dilution Module and a Filtration Module, with a separate Sample Collector (collectors ASX-7200 and ASX-7400 are named). It is coordinated by Lucullus software, which the vendor describes with sample planning, barcode and label printing, automatic data import and feedback control [S1, S2].

QB AutoSampler is a modular, software-defined benchtop sampling system. A QB Syringe Pump and a QB Selector Valve form a sampling path, a QB Edge powers and controls the modules over PoE Ethernet, and QB Control provides the sampling services and the recipe engine. Cooled storage adds the QB Sample Cooling Chamber with a QB Thermo chiller [QB-K, QB-C].

Sources, destinations and analysers

Numera documents sampling from up to 16 bioreactors (up to four Multiplexer Modules, each taking up to four bioreactors or other sampling sources) and an interface to Ambr 250. Its Routing Module provides direct transfer to a maximum of six analysers, and direct HPLC injection valves rated up to 20,000 psi are listed. A supported-analyser datasheet (04/2026) lists Vi-CELL XR and BLU, Nova FLEX2, Roche Cedex Bio HT and HiRes, Agilent 1200 Infinity, and Thermo UltiMate 3000 and Vanquish as compatible analysers; the mode of transfer is not stated [S1, S2, S15].

A 14 October 2026 press report states that Securecell and Beckman Coulter Life Sciences are building an automated Numera to Vi-CELL BLU transfer with Lucullus integration. It is in development, with release expected in about nine months, so it is announced and not available [S16].

In the QB AutoSampler, one Selector Valve offers up to 24 positions, which are shared between source, wash, air, waste and destination ports; valves are stackable. Paths can be shared (one pump and valve pair serves several sources), dedicated or combined. Vials, analytical devices and waste are destination types, but QB documents no specific analyser integration and no maximum number of analysers. Because valve positions are not sources, the two systems’ source counts are not compared number to number. QB Systems has not published a list of verified bioreactor interfaces, so confirm port compatibility per case [QB-C, QB-K, QB-O].

Sample preparation

Securecell NumeraQB AutoSampler
DilutionDilution Module, 1:2 to 1:100; precision not stated in the catalogue; mechanism not described on the product page [S1, S2]In-vial, dilution factor 1.0 to 6.0 (total volume divided by sample volume), air chase on by default; the instrument does not mix; no numeric ratio accuracy specified; sample at least 1 mL when diluting; one destination per run [QB-C]
SplittingOne sample to up to six analysers (Routing Module) [S1, S2]Sampling Aliquoting: one draw to two ports, one after the other; operator-entered dead volumes; no numeric split accuracy specified [QB-C]
FiltrationMembrane-based tangential filtration, PES tapes of 0.22, 0.45 and 1.20 µm; up to 375 filtrations per coil (product page), 30 m coils (catalogue) [S1, S2]Not documented

Vendors do not define their ratio notation on the pages reviewed (for example whether 1:2 means one part sample to one or to two parts diluent), so QB’s factor D and the Numera ratios must not be compared number to number.

Sample volume, cleaning and storage

Securecell NumeraQB AutoSampler
Sample volumeFrom 2.7 mL (product page; the catalogue does not state it) [S1]1.0 to 6 mL recommended (6 mL needs the 10 mL syringe); dead volume 0.5 to 1.0 mL; path accuracy approx. ±0.5 mL [QB-O, QB-C]
CleaningEthanol rinsing followed by drying of the line; air blow-back, pinch-valve barriers (described by the vendor as sterility barriers) and a collector rinse station; autoclavable sampling sets [S1]Configurable wash, rinse and air-dry legs around each sample, optional path fill, a port-level Sampler Cleaning service; dedicated consumable sets, an A4BEE-provided SOP and an aseptic testing report on request [QB-C, QB-O]
StoragePeltier racks for 1.5, 4.0 and 10 mL vials (40, 24 and 12 vials per rack) and deep-well plates (2, 5 and 10 mL); 4 to 20 °C (product page) and 4 to 40 °C at 25 °C ambient (catalogue) [S1, S2]QB Sample Cooling Chamber: 42 × 4 mL, 36 × 8 mL or 21 × 4 mL plus 18 × 8 mL vials; 4 °C for more than 72 hours, only while the Temperature Control by TCU service is in Execute; 1 °C minimum reached in laboratory conditions [QB-K, QB-O, QB-C]

Triggers, control and integration

Numera documents sampling “as often as every 6 minutes”, 24/7. Triggers are not stated on the pages reviewed. The Lucullus software adds sample planning, automatic data import and feedback control [S1].

The QB AutoSampler runs scheduled sampling and conditional sampling built in the QB Control Recipe Designer, with Wait Until conditions on live device tags, timers, counters, variables and formulas. The OPC UA connector can run an OPC UA server for external clients, and the services publish status and lifecycle events for external clients. Third-party control systems can trigger sampling through the OPC UA connector (owner statement; no sampling-specific mapping has been found in the code, so confirm the setup with QB Systems). This comparison makes no claim about closed-loop control driven by analyser results; QB documents no such function [QB-C, QB-O].

Data integrity and licensing

The Numera pages reviewed state no 21 CFR Part 11 position and no licence terms [S1, S2].

QB Control records an audit trail and provides PDF batch reports, including a Batch Run Audit Trail, a Recipe Audit Trail and a “Sampling basic” report for batches that ran Automated Sampling. QB Control provides 21 CFR Part 11 conformance features (audit trail, PDF reports); electronic signatures are not yet covered (in development, no committed date), so labs that need them must cover them in a separate system. This is not a claim of compliance or validation, which is a property of the user’s validated process. The QB AutoSampler is positioned as non-regulated research and process-development tooling. QB Control is licensed per hardware modules and services in the tiers BASIC, ADVANCED and PRO, perpetual or fixed-term; scaling with new hardware needs a licence upgrade. This page does not compare commercial amounts [QB-C, QB-O].

Where the QB AutoSampler has limits

  • Dilution is in-vial only, up to a factor of 6, without mixing and without a specified ratio accuracy.
  • Splitting a draw goes to two ports, one after the other; Numera publishes distribution to up to six analysers.
  • Analyser transfer. QB documents no specific analyser integration and no maximum number of analysers; Numera documents a named list of supported analysers.
  • Filtration is not documented for the QB AutoSampler.
  • Volumes. Dead volume of 0.5 to 1.0 mL per sample and a path accuracy of approx. ±0.5 mL are disclosed; comparable figures were not found in the Numera public sources, so no conclusion is drawn either way.
  • Cooling continuity. Cooling runs only while the Temperature Control by TCU service is in Execute, so it stops when the service is paused, held, stopped or restarted, or when the message bus connection is lost.
  • Positioning. QB Systems is a newer entrant in dedicated bioprocess sampling, and this comparison cites no application data for the QB AutoSampler.

Making the decision

Ask each vendor the same questions. If samples must reach several at-line analysers directly, compare the documented analyser counts and integrations. If you need wide dilution ranges or filtration inside the instrument, Numera documents them. If a small in-vial dilution factor is enough and you want sampling, cooled storage and bioreactor control in one platform with recipes and an audit trail, the QB AutoSampler may fit. The full analysis, which also covers Eppendorf, Flownamics and Merck, is in the article The State of Benchtop Automated Sampling in Bioprocessing.

Sources

Evidence labels for QB statements: QB-C = QB Systems documentation and platform behaviour (state of 14 October 2026); QB-K = QB product documentation 09.2026; QB-O = statements of QB Systems (product owner), not verified in the product code. Competitor sources were retrieved on 19 October 2026.

  1. S1 — Securecell, Numera product page
  2. S2 — Securecell, Numera Product Catalogue 2026
  3. S15 — Securecell, Numera supported analytical devices datasheet (04/2026)
  4. S16 — Bionity, Securecell and Beckman Coulter Life Sciences team up (press report, 14 October 2026; vendor announcement)

Analysis prepared October 2026 | QB AutoSampler vs. Securecell Numera comparison QB Systems — Flexible. Innovative. Limitless. | A product brand of A4BEE Sp. z o.o. | https://qb.systems